Background: Numerous laboratories in Italy use radioimmunoassay to determine concentrations of sex hormones (FSH, LH, testosterone). A comparison of assay methods is thus an important starting point for the achievement of universally accepted reference values. Aim: To carry out an external quality assessment for FSH, LH, and testosterone. Materials and methods: Fifteen aliquots from 5 serum pools were assayed in multiple replicates by 16 Italian laboratories with 5 automated immunoassays (Abbott Architect, DiaSorin Liaison, Perkin-Elmer AutoDelfia, Roche Elecsys, Siemens Immulite 2000), and 1 radioimmunoassay (Adaltis). Results: The variance was below 12% for FSH, between 11.61% and 14.76% for LH, and between 9.57% and 12.48% for testosterone. Assay precision was good, except for Elecsys at low concentrations of FSH and for Immulite at low concentrations of LH and testosterone. ARCHITECT showed a negative bias for FSH and LH and a positive bias for testosterone; Liaison a positive bias for LH; Elecsys a positive bias for FSH and a negative bias for testosterone; Immulite a positive bias for FSH; AutoDelfia a negative bias for FSH and a positive bias for testosterone. Reference ranges at the low end varied widely, even among laboratories using the same assay. Conclusions: The analytical performances of widely used immunoassays for FSH, LH, and testosterone show a fair to strong degree of consistency. A careful evaluation of reference ranges by clinical and laboratory experts needs to be carried out, in order to reach a consensus. ©2013, Editrice Kurtis.

A multicenter evaluation of immunoassays for follicle-stimulating hormone, luteinizing hormone and testosterone: Concordance, imprecision and reference values

Lenzi, A.;Spaziani, M.;
2013-01-01

Abstract

Background: Numerous laboratories in Italy use radioimmunoassay to determine concentrations of sex hormones (FSH, LH, testosterone). A comparison of assay methods is thus an important starting point for the achievement of universally accepted reference values. Aim: To carry out an external quality assessment for FSH, LH, and testosterone. Materials and methods: Fifteen aliquots from 5 serum pools were assayed in multiple replicates by 16 Italian laboratories with 5 automated immunoassays (Abbott Architect, DiaSorin Liaison, Perkin-Elmer AutoDelfia, Roche Elecsys, Siemens Immulite 2000), and 1 radioimmunoassay (Adaltis). Results: The variance was below 12% for FSH, between 11.61% and 14.76% for LH, and between 9.57% and 12.48% for testosterone. Assay precision was good, except for Elecsys at low concentrations of FSH and for Immulite at low concentrations of LH and testosterone. ARCHITECT showed a negative bias for FSH and LH and a positive bias for testosterone; Liaison a positive bias for LH; Elecsys a positive bias for FSH and a negative bias for testosterone; Immulite a positive bias for FSH; AutoDelfia a negative bias for FSH and a positive bias for testosterone. Reference ranges at the low end varied widely, even among laboratories using the same assay. Conclusions: The analytical performances of widely used immunoassays for FSH, LH, and testosterone show a fair to strong degree of consistency. A careful evaluation of reference ranges by clinical and laboratory experts needs to be carried out, in order to reach a consensus. ©2013, Editrice Kurtis.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11389/62799
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 22
  • ???jsp.display-item.citation.isi??? ND
social impact